iPRO 5000 Series TN/TS Analyzer Installation Procedure
Page 26 of 37 Version 1.2
Install the 105- or 8-position tray. Make sure that it is level with the sampler base casting.
Fig 5.8 Placing the tray
Section completed:
6.5.5 Placing the needle
Open the transparent door of the AS3000 II Autosampler to reach the sampler needle. Make
sure the needle holder is well positioned: the top should be placed flush with the housing (see
red circle in the pictures below) and the black lock (displayed lower, under the red circle)
should be in position.
Fig 5.9 Incorrect installed needle (left) and correctly installed needle (right)
Place the AS3000 II Autosampler needle in the needle holder. Lift the needle holder in order
to guide the tip of the needle through the hole in the needle holder.