iPRO 5000 Series TN/TS Analyzer Installation Procedure
Page 23 of 37 Version 1.2
6.5 AI/AS3000 II Autosampler
Check which sampler tray is used and take the appropriate positioning tray. The delivery of
the tray depends on the order placed. There are two possibilities; both are shown in the
pictures below, the 105-position tray (top) and the 8-position tray (bottom).
Fig 5.2 Positioning tray with and without the corresponding sample tray
Section completed:
6.5.1 Thermal Switches
The Thermal Switches may be activated due to transportation. Press the thermal switches
located at the left side of the centring plate in fully for 2-3 seconds.
Fig 5.3 Thermal Switches overview
Section completed: